How do you calibrate a VNA?

From RF Design Wiki · By the · Published · Last reviewed · How we research

VNA calibration measures known standards to solve for the systematic error terms of the instrument and cables, then removes them mathematically. The classic two-port model has 12 error terms (directivity, source match, reflection tracking, load match, transmission tracking and isolation in each direction). Common methods are SOLT (short-open-load-thru), TRL (thru-reflect-line) and automatic electronic calibration modules.[1][2]

When should SOLT or TRL be used?

What mistakes spoil calibration?

How is a calibration verified?

Measure a verification standard (an airline, a mismatched airline such as a Beatty standard, or attenuators) and compare with its known data. A quick check: a calibrated load should read a return loss well beyond 40 dB, and a thru should show near 0 dB with flat phase after delay removal.[1]

Frequently asked questions

What is the 12-term model?

Six error terms per direction for a two-port VNA with three receivers.

How often should I calibrate?

Whenever the setup changes, and periodically as temperature drifts.

Does calibration remove fixture effects?

Only if the calibration plane is at the DUT; otherwise de-embed.